Temperature Validation System Q-TAS
Q-TAS Thermocycler Validation
The Q-TAS system is a powerful temperature validation system for thermal cyclers. It includes a control module with an integrated real-time display, probe plates for fixed or variable probes, and powerful PC software.
Q-TAS reliably analyzes data and automatically generates reports. For maximum flexibility and mobility, the system features eight variable probes.
Request a quoteQ-TAS System
Key components include a control module with integrated real-time display, probe plates that accommodate either fixed or variable probes, and powerful PC software. Despite being extremely powerful, the TAS-System is also very easy to use. Allow the TAS-Software to analyse the data for you and to automatically generate a detailed report for each test.
Specifications
- Simple, quick operation: Measure and analyse thermal cycler temperature performance in under 10 minutes, the TAS-System allows for visual comparison from test to test and enables performance tracking over the lifetime of the thermal cycler.
- Flexibility: As well as its standard fixed probe plate, the TAS-System is offered with a variable probe plate option. This utilises individually interchangeable temperature probes that can be placed in any of its 96 well positions. Combined with leaded-probes, the variable probe plate offers yet further flexibility, becoming the ideal solution for testing of non-standard thermal cyclers.
- Measurement integrity: Each interchangeable TAS probe is uniquely identified, allowing for the automatic detection of probe position, the application of specific calibration data for each probe, and a calibration expiry warning when relevant.
Dimensions (W x D x H) | 125 x 90 x 40 mm |
Weight | 350 g |
Measurement range | 0°C to 100°C |
Number of temperature probes | 8, 16, 24, 48, 96 (je nach Modell) |
Detection of sensor positions | automatic |
Response time of the temperature sensor | < 0.2 seconds |
Probe temperature | Specific calibration data for each probe |
Temperature accuracy | ± 0.1°C @ 95 °C, QTAS system PC controled ± 0.2°C @ 25 °C, QTAS system Stand-alone operation ± 0.1°C @ 95 °C, after download of data on PC |
Calibration interval | 12 months |
Operating system | Windows 2000, XP, 7, 10 |