Laboratory UV Irradiation System Bio-Sun
BioSun is a complete, microprocessor-controlled UV irradiation system for Petri dishes and microplates.
The gold standard for UV irradiation of cells.
Based on a programmable microprocessor, the system constantly monitors the UV light emission.
The irradiation stops automatically when the energy received matches the programmed energy.
Thanks to its advanced UV sensors, the irradiation cycles are perfectly reproducible, regardless of the intensity fluctuation of the UV source.
Request a quoteBioSun Irradiation System
Spezifications
- Irradiation modes:
312 nm, 365 nm or 312 / 365 nm combined, dose-oriented programmable - UVC 254 nm germicidal lamps fo tray sterilisation
- Fully microprocessor controlled, thus perfectly reproducible irradiation results
- UV safety shutdown when opening the sample chamber
- Pull-out, large sample drawer with stainless steel holder
- Extremely homogeneous UV illumination
- Temperature control: maximum temperature of 30°C in the radiation area at 20°C room temperature
- Calibrated UV sensors
- Continuous acquisition of UV sensor data, 4 measurements per second
- Display of time, UV intensity, temperature and energy
- Epoxy-coated aluminium housing
Models
- BioSun ++:
The TOP system for controlled UV irradiation of cultured cells - BioSun UV.H:
The gold standard for controlled UV irradiationof cultured cells
Option
Optional equipment USB 2.0 interface with control software:
Programming and control via PC
graphic recording of irradiation processes
Database structure for data storage.
Benefits
- Large irradiation area, homogeneously illuminated
- Continuous control of the UV irradiation and data acquisition by calibrated UV sensors, re-calibration possible with certificate
- Irradiation time less than 40 seconds for a programmed dosis (energy) of 150 mJ/cm²
- Up to 6 UV measurements per second
- Tray temperature does not exceed 30°C (86°F) for room temperature of 20°C (68°F)
- Optional interface enabling connection between the BioSun and a computer for data monitoring and recording
General Technical Data BioSun
Model | BioSun ++ | BioSun UV.H |
---|---|---|
Irradiation area for Petri dishes | 260 x 120 mm | 260 x 120 mm |
Irradiation area for microtiter plates | 260 x 90 mm | 260 x 90 mm |
UV sources | 4 x 30 Watt, 365 nm 2 x 30 Watt, 312 nm | 4 x 30 Watt, 365 nm 1 x 30 Watt, 312 nm |
Irradiation homogeneity | 312 / 365 nm: ± 7 / 7% | 312 / 365 nm: ± 5 / 7% |
UV sterilization of the irradiation area | 1 x 30 Watt, 254 nm tube | 2 x 30 Watt, 254 nm tube |
Irradiation distance | 25 mm | 25 mm |
BioSun system dimensions | 1.100 x 500 x 600 mm | 1.100 x 500 x 600 mm |
Voltage | 230 V | 230 V |
Order No. | 110.0172 | on request |