Electrophoresis Power Supply

The most robust, long lasting and durable electrophoresis power supplies in the market.

The manual mode permits to set voltage, current, power and time for a routine electrophoresis run.
Parameters can be changed temporarily without interrupting the run.

*For the exact Order No. refer to the Technical Data.

Request a quote

Power Supplies for Electrophoresis

Method programming 
Up to 9 different programs, each with 9 steps, of frequently used parameters can be stored in the non-volatile memory for future recall. Reads voltage in 1 V steps, current in 1 mA steps and power in 1 W steps. Each step is able to recall a next one, providing a flexible multiple step function for special techniques. Parameters of the running step can be changed temporarily without interrupting the run.

Voltage ramp 
The method mode also permits to program a linear voltage gradient for any step provided the limiting current or power is not attained.

Timer 
Timer or volt-hour controlled operation is a useful standard feature on all models. The microcomputer will automatically terminate the run when the count down of the selected value is achieved.

Automatic cross-over 
Each model has constant voltage, constant current, constant power capabilities with automatic cross-over and shows which parameter is kept constant.

Automatic recovery after power failure 
In case of a mains drop out the instrument will automatically continue the run for the remaining time.

Data-logging
Stores periodically up to 3600 output values (voltage, current and power, time or volthours) including program number and step.

Data transfer 
A free data acquisition software for PC is available optionally. Download software (attention: .NET Framework 1.1 is required!). It permits to visualise and examine the stored run details.

Remote control 
All power supplies can be controlled by a computer using special commands.

Safety features
The user is protected from potential shock hazard since the AC line is automatically disconnected from the high voltage transformer when a ground leakage path is detected. The instrument is fully protected against any overload condition including accidental short circuit of the output. The high voltage cannot suddenly appear at the outputs. It will always increase smoothly up to one of the pre-set limits is reached. Galvanic RS232 input/output insulation prevents dangerous ground loop interferences when connected to a computer. (Model EV222 has no method programming, data-logging or remote control possibility).

150 Watt

ModelVoltage
0 to ...
Current
0 to ...
Power
0 to ...
Number of Socketsmin. Load ResistanceDimensionsWeightOrder No.
EV 2310300 V150 W1000 mA4 x 410 Ω24 x 20 x 13 cm6 kg101.1053
EV 2650600 V150 W500 mA4 x 430 Ω24 x 20 x 13 cm6 kg101.1055
EV 22301500 V150 W300 mA4 x 4300 Ω24 x 20 x 13 cm6 kg101.1052
EV 23203000 V150 W150 mA4 x 4600 Ω24 x 20 x 13 cm6 kg101.1054

300 Watt

ModelVoltage
0 to ...
Current
0 to ...
Power
0 to ...
Number of Socketsmin. Load ResistanceDimensionsWeightOrder No.
EV 3020300 V300 W2000 mA4 x 45 Ω31 x 26 x 13 cm10 kg101.1056
EV 3610600 V300 W1000 mA4 x 415 Ω31 x 26 x 13 cm10 kg101.1059
EV 31501200 V300 W500 mA4 x 470 Ω31 x 26 x 13 cm10 kg101.1057
EV 33303000 V300 W300 mA4 x 4600 Ω31 x 26 x 13 cm10 kg101.1058
EV 36206000 V300 W150 mA4 x 21200 Ω31 x 26 x 13 cm10 kg101.1060

 

Add-ons

Agarose Gel Chamber - Mini
Horizontal Gel Systems
Agarose Gel Chamber - Midi
Horizontal Gel Systems
Agarose Gel Chamber - Midi Large
Horizontal Gel Systems