Designed for clean operations with DNA samples.
Equiped with a flow-type bactericidal UV cleaner-recirculator AR, which provides constant decontamination inside the cabinet during operation.
This model is also available as a large version.
Dimensions (WxHxD) | 690 x 555 x 585 mm |
---|---|
Part Number: 112.0014
UV-radiation from the open lamps disinfects the working area inactivating DNA/RNA fragments during 15–30 min of exposure. A digital timer controls duration of the direct UV irradiation. A daylight lamp provides proper illumination of the working surface.
UV cleaner-recirculator AR consists of a UV lamp, a fan and dust filters organized in a special body so that a user working with a UV-cabinet is protected against UV light. Recirculator increases the maximum density of UV light making it sufficiently effective for DNA/RNA inactivation. The UV-recirculator processes 360 UV-cabinet volumes per hour, creating permanent aseptic conditions of operation inside the UV-cabinet
Advantages
Dimensions: | 690 x 555 x 555 mm |
Working area: | 650 x 475 mm |
Weight (net / gross): | 28.8 / 39 kg |
Open UV lamp: | 1 x 25 W, bactericidal |
UV radiation level | 15 mW / cm2 / sec |
Radiation type: | Ultraviolet (λ = 253,7 nm) ozone-free |
Digital time setting of direct UV exposure | 1 min - 24 h / non-stop (increment 1 min) |
UV-recirculator: | 1 x 25 W (efficiency > 99 % per 1 hour) |
Daylight lamp: | 1 x TLD-15W (for working area illumination) |
Glass type: | Euroglass (Germany) |
Thickness of glass: | sides = 4 mm Upper front panel = 8 mm Screen = 4 mm |
UV protection film type: | 4 MIL Clear |
Optical transmission: | 95 % |
UV protection: | > 96 % |
Safety features: | Automatic open UV-lamp switch off when screen is open |
LED time controller of UV exposure: | 0 - 24 hours |
Power: | 240 V, 50/60 Hz |