PCR Thermal Cycler TurboLite
TurboLite is a powerful, reliable and affordable thermal cycler for all standard PCR applications.
It has gradient function that enables the user to set maximum 24 °C temperature gradient between the left and right columns in the sample block at each temperature.
TurboLite has a high heating rate, fast boot up, and accurate temperature control. It ensures efficient and accurate PCR processing.
Applications
- PCR
- CRISPR assays
- NGS sample pretreatment
Gradient Thermocycler TurboLite
The unit is compact in size and has front-to-back venting which allows machines to be placed side-by-side.
This feature helps to reduce heat interference with other equipments.
All these features make it an ideal solution for growing service centers and the education market while keeping the cost down.
Temperature increment/decrement function makes touchdown PCR possible.
Unique design enables slow ramp incubaion for CRISPR/Cas related assay.
- Intuitive operation experience
Touch keypad + intuitive graphic interface
The large liquid crystal module (LCM) display with touch keypad enhances visibility and ease-of-operation. - Easy to use lid construction
The lid has an adjustable heating temperature. The pressure loaded lid prevents the formation of condensation. - Fully adjustable lid temperature
The temperature can be set between 35 and 120°C for virtually any type of experiment including NGS pre-treatment. - Advanced slow-ramp temperature control
The ramp rate can be precisely controlled down to 0.1 °C/sec to meet the need for the CRISPR/Cas related assays - A simple and easy-to-use graphical interface
- Gradient optimization
The gradient function enables optimization of annealing temperatures - Preset protocol templates make starting amplification easier
- Automatically creates operating history and error logs
- Easy disinfection
The dust and aerosol proof keypad can be easily disinfected - Auto restart
Power failure recovery keeps the experiment safe
Order No. | 112.6000 |
Sample Block | |
Sample Block | Fixed 96-well, compatible with regular profile or low profile 0.2 ml PCR tube, strip, non-skirted, semi-skirted and full-skirted 96-well plate |
Sample Volume | 10 – 100 µl |
Block Temperature | |
Temperature Range | 4.0 – 99.9 °C |
Max. Heating Rate | 3.0 °C/sec |
Max. Cooling Rate | 2.0 °C/sec |
Temperature Accuracy | ± 0.3 °C |
Temperature Uniformity Across Block | ± 0.5 °C |
Adjustable Ramp Rate | High / Low |
Slow-Ramp Temperature Control | Via temperature increment / decrement between cycles |
Gradient Temperature | |
Gradient Direction | Horizontal across the block |
Gradient Temperature Range | 30 – 99 °C |
Gradient Temperature Difference | Max. span 24 °C |
Heated Lid | |
Temperature Setting Range | 35 – 120 °C or off |
Temperature Accuracy | ± 1.0 °C |
Software | |
Stored Program No. | 200 sets |
Registered User Folder No. | 50 sets |
User Folder Password Protection | Yes |
General | |
Display | Monochrome LCM, 240 x 128 pixels |
Input Interface | Capacitive touch keypad |
Auto Restart after Power Outage | Yes |
Footprint Dimensions (H x W x D) | 225 x 245 x 415 mm |
Weight | 9.0 kg |
Volume (Measuring distance approx. 40 cm) | Heating: ~ 47 dB Cooling: ~ 57 dB |
Power Supply | VAC 240 V, 50/60 Hz, 500W |
Certification | CE, RoHS |