Microcentrifuge MSC-2P

The combination of centrifugal and vortex functions ensures fast operation, thorough mixing, and consistent results.
The MSC-2P can accommodate two PCR plates simultaneously (with or without rims), resulting in significant time savings in the PCR workflow.
Applications of MSC-2P
- Gen expression analysis and genotyping
- Cloning
- Mutagenesis
- Methylation analysis
- Sequencing
- Medical, Forensic, and Applied Sciences sample and reagent
The plate centrifuge was specifically designed to achieve improved PCR yield in subsequent analyses. This is accomplished by collecting droplets, mixing reagents, and recollecting the samples.
Specifications
- Spin-Mix-Spin algorithm (SMS)*
Collects micro-volumes to the bottom of PCR plates with repeated centrifuge-vortex cycles. - Customizable SMS cycles
1–10 cycles for optimal results - Dual functionality
Centrifuge mode: for fast spin-down vortex mode full rotor shake for uniform mixing - High speed performance
Max. speed: 3.500 rpm
Max. RCF: 610 × g
Four independent modes
- Centrifuge: max. 3500 rpm
- Vortex: up to 5 min.
- Centrifuge/Vortex: combined two motion types
- Spin-Mix-Spin algorithm: up to 10 cycles
Compatible Plates
- Semi-skirted 96-well plates
- up to 8–8.5 mm (use adapter set MSC-2P-1 or
adapter set MSC-2P-2 provided with device)
- up to 9–10 mm high. - Unskirted 96-well plates
- 1.2 mm high (use adapter set MSC-2P-1)
- 2 mm high (use adapter set MSC-2P-2. - 384-well plates up to 10 mm
- PCR strip tubes using optional SH-12x8 adapter
| Rotations specifications | |
| Capacity | 2 PCR mikroplates |
| Zentrifugationsmodus | 500 – 3500 rpm |
| Rel. centrifugal force | up to 610 g |
| Centrifugation timer | 1 s – 10 min, step 1s |
| Vortex-Modus | Full rotor |
| Timer for vortex | 1 – 5 min, step 1 s |
| SMS cycle regulation | 1 – 10 cycles |
| Max. operation time | 150 min |
| General technical data | |
| Tube protection | Lid lock |
| Display | LCD, 16 x 2 signs |
| Temperature range RT | +4 to +40°C |
| Max. humidity | 80% (at 31°C RT) |
| Workroom description | Cold rooms, incubators (except CO2 incubators) and closed laboratory rooms |
| Dimensions (W x D x H) | 165 x 220 x 230 mm |
| Weight | 2.7 kg |
| Power consumption | 18 W |
| Power supply | Input AC 100-240 V 50/60 Hz, Output DC 12 V, 2 A |