Circulating Coolers ULK

The ULK series comprises high-performance, space-saving, and quiet recirculating chillers for laboratories, production facilities, and industrial applications. Customizable through a wide range of available options.

Environmentally friendly and future-orientated thanks to natural refrigerants: reduced refrigerant quantity and considerably lower GWP values (global warming potential), as well as low power consumption.

Please refer to the technical data for the different models and the exact article number.

Request a quote

Circulating Chiller ULK Series

Characteristics

  • Environmentally friendly
    - energy-saving fan with EC technology
    - natural refrigerant with high efficiency reduces power consumption    
    - lower refrigerant charge quantities per device
  • Low operating noise
    - due to particularly quiet components
    - fan adjusts the speed to the demand
  • User-friendly
    - easy filling and emptying thanks to integrated filling funnel and drain valve
    - leak-free hose replacement thanks to self-sealing hose connector with quick coupling
  • High-quality components
    - Touchscreen controller with glass surface and integrated digital display for fill level and pump status
    - Brand-name components for long service life
    - Hermetically sealed, air-cooled, and low-maintenance refrigeration unit
  • Operational safety
    -  Freeze and overheat protection
    - Automatic pump shut-off in case of low coolant level
    - Visual and audible alarms
    - Clear-text error messages for quick diagnostics

Options

  • RS485 interface for MODBUS communication: Enables integration into control systems.
  • RS485 interface for ST-Bus communication: Allows easy reading and modification of values.
  • Voltage input for external setpoint control: Allows direct control of the temperature via external input signals.
  • Voltage output for actual value transmission: Transmits the current temperature values to an external system.
  • Potential-free contact: Enables connection to an external alarm system for enhanced safety.
  • Direct temperature measurement in your application: An external sensor measures the temperature directly at the process point. Automatic adjustment of the setpoint ensures particularly stable temperature control and greater process reliability compared to standard systems.
  • Mobile version: four castors, two lockable.

Accessories

  • Thermo-insulated Hoses:
    - Hose with thermal insulation and hose clamps
    - Ø 10mm: for hose connection Ø 10 mm
    - Ø 12,7mm: for hose connection Ø 12-13 mm
    - Down to -20°C, 16 bar, PVC, reinforced
    - Sold by the metre
  • HKF coolant 15.1 POF ECO:
    - Recommended for all FRYKA circulating coolers
    - 5 litres container
    - Ready-made mixture
    - Frost-proof up to -15°C
    - Incl. practical dispensing tap
ModelULK 602 NULK 1002 NULK 2002 N
Order No.104.1501 N104.1504 N104.1505 N
Operation temperature range-10°C to +40°C-10°C to +40°C-10°C to +40°C
Cooling accuracy at -10°C+/- 0.5 K+/- 0.5 K+/- 1 K
Display resolution0.1 K0.1 K0.1 K
Cooling Capacity
at +20 °C600 W1000 W2000 W
at +10 °C500 W800 W1550 W
at 0°C350 W550 W900 W
at -10 °C200 W300 W650 W
Circulation Pump
Flow rate (max.)10 l/min12 l/min12 l/min
Flow pressure (max.)0.6 bar2.9 bar2.9 bar
General Technical Data
External dimensions (W x D x H)310 x 415 x 585 mm350 x 450 x 680 mm400 x 450 x 730 mm
Installation dimensions (W x D)310 x 715 mm350 x 750 mm600 x 750 mm
Weight28 kg39 kg42 kg
Noise level (distance 1 m)49 dB(A)52 dB(A)52 dB(A)
Ambient temperature range+5°C to +40°C+5°C to +40°C+5°C to +40°C
Max. permissible temperature of the coolant60°C60°C60°C
Electrical connection230 V / 50 Hz230 V / 50 Hz230 V / 50 Hz
Power consumption380 W800 W950 W
Max. current 3.0 A5.0 A6.5 A
Coolant tank1.5 to 4.0 litres2.5 to 9.0 litres2.5 to 9.0 litres
Hose connections12.7 mm12.7 mm12.7 mm
Protection classIP 20IP 20IP 20
Class. acc. DIN 12876I / NFLI / NFLI / NFL
Max. operating pressure25 bar25 bar25 bar
RefrigerantR290R290R290
Minimum room volume12 m312 m312 m3