RT Cycler MyGo ESR

The MyGo Pro ESR cyclers are multiplex real-time PCR devices suitable for a wide range of applications including life science research, food speciation testing and virus detection and quantification.

MyGo ESR is the extended version of MyGo Pro and offers advantages in the red channel area.

Consumables for all MyGo models

Request a quote

MyGo ESR - Real-Time PCR Cycler

Powerful Peltier elements and solid silver blocks ensure both speed and very stable thermal uniformity. The result is fast, accurate, quantitative PCR and melting point analysis.

Advanced algorithms combined with an intuitive user interface support a broad range of applications, operating systems and connection options.

MyGo ESR versus MyGo Pro

1. the automatic quantification analysis is up to three times faster
2. quality values are more robust over a wider range of signal-to-noise ratios
3. multiple wells can be included or excluded together in results tables

These top features remain

1. no moving parts for absolute transport safety
2. independence from consumables
3. optical and thermal validation possible through MyGo validation kit: important for all accredited laboratories

Faster Runs
You can now programme faster PCR runs. Previously, hold times were a minimum of 10s, followed by optical capture.
In most cases, hold times can now be a minimum of 3 seconds, including hold times with optical captures for the MyGo Mini.

Hold Time MyGo ESR

Hold time with detectionHold time without detection
17 s3 s
Optical Data
No. of channels5 (Multiplex)
Fluorescence Excitation610 nm and 450 nm (red/blue LED)
Fluorescence Detection510 nm to 750 nm
Detector typeCMOS array
Detection sensitivitySingle copy detection
Dynamic range9-log
Presicion1.1 fold discrimination
Temperature Control
Thermal systemPeltier, 99.99% silver block
Temperature range37°C to 99°C
Speed5°C/s Heating
4°C/s Cooling
Thermal uniformity± 0,1 °C
Thermal accuracy± 0,25°C
Factory calibrated dyesSYBR Green I, ResoLight, FAM, VIC, HEX, Cy5, ROX, LC Green, SYTO 9, Cy5.5. Instrument Specific Calibration for (FAM, HEX, ROX, CY5, CY5.5) Note: Does not work with TexasRed.
Multiplex with red dyesYes
Supported assay formatsTaqMan®, Molecular Beacons, SimpleProbes®, Intercalators,
High Resolution Melting (HRM)
General Technical Data
Order No. MyGo Pro ESR112.2015
Number of wells32 well
Tube format0.1 ml tubes, 8 well strip-tubes
Reaction volume10 – 100 μl
Run time<40 min (dependent on assay)
Status displayLighting in the lid changes color
InterfaceDesktop PC/laptop required
Note: Protocols can be run from
a preloaded USB thumb drive.
SoftwareWindows/Mac OS X or Linux operating software
ConnectivityLAN (RJ45), USB
Dimensions (W x D x H)25 x 27 x 23 cm
Power consumption170 W
Weight7 kg
Electrical Voltage, Frequency100 – 240 V AC +/–10%,
50 – 60 Hz +/–10%
Operation noise<40 dB