EzDrop 1000C Spectrophotometer

With EzDrop 1000C spectrophotometer, measurements of sample concentrations can be made within three seconds.

The unit offers two modes for microvolume and cuvette detection, as well as an intuitive touchscreen for easy operation. These features meet a variety of nucleic acid and protein quantification needs.

Specifications EzDrop 1000C

  • Gold-standard precision at a competitive price
  • 3-second measurement for DNA, RNA, and protein
  • Works standalone or connects to PC
  • Dual system for microvolumes and cuvettes
  • Smart pipetting assist light for accurate placement
Request a quote

Spectrophotometer for Cuvette and Microvolumes

  • Independent system with clear touchpad
  • Multiple automated measurement functions allow you to quickly measure nucleic acids, proteins, cells, and other samples at the touch of a button.
  • Sophisticated design: an additional light source compensates for the lack of ambient light and minimizes errors in sample placement
    (LED Assist Light).
  • A special nano-hydrophobic layer on the sample window ensures the formation of a stable sample column.
  • The sample glass can be completely replaced, significantly reducing the possibility of residual contamination
  • CV absorbance value is <1%.
  • The special design of the detection arm reduces pressure, which in turn reduces the error rate in experiments.
  • Data exchange: USB connection (csv-file)

Application Areas

  • Nucleic acid and protein analysis
  • OD600 measurement
  • Analysis of microarrays and labeled proteins
  • KineticsFull spectrum scans

Cuvette Holder

Microvolume pickup with LED assist light
Sample window with nanohydrober coating ensures stable sample column.

General Technical Data EzDrop 1000C

Order No.
EzDrop 1000C
103.3910
Display7” touch screen, 1280*800 high-definition color display
ConnectivityUSB-A port x1 (Data output, csv-file); USB-B port x1 (PC connection)
Footprint dimensions
(W x D x H)
206 x 333 x 166 mm
Weight3.3 kg (7.3 lbs.)
Glove suitabilitySuitable for all laboratory gloves
Internal storage32 GB flash memory
Operating voltageInput: AC 100-240 V, 50/60 Hz;
Output: DC 24 V, 2.08 A

 

Specifications EzDrop 1000C

Microvolume Mode
Minimum sample volume1 µl
Sample number1
Pathlength0.5 mm / 0.05 mm
Light sourcePulsed xenon flash lamp
Detector type2048 element CMOS
Wavelength range190 - 1000 nm
Wavelength accuracy1.0 nm
Spectral resolution1.5 nm (FWHM bei Hg 253.7nm)
Absorbance precisionRaw: 0.0015 A (0.5 mm);
0.03 A (1 cm equivalent)
Absorbance accuracy1.5 % bei 1.0 A and 320 nm
Absorptionsbereich (entspricht 1 cm)0.04 - 400 A
Detection rangedsDNA: 2 – 20000 ng/µl
BSA: 0.06 – 600 mg/mL
Measurement time< 3 sec.
Cuvette Mode
Pathlength10, 5, 2, 1, 0.5, 0.2, 0.125, 0.1 mm
Absorbance range (1 cm equivalent)0.002 - 1.5 A
Detection rangedsDNA: 0.3 - 75 ng/µl;
BSA: 0.003 - 2.25 mg/ml
Stirring8 speeds
150 – 850 rpm
Temperature control37 - 45°C  ±0.5°C
(Quartz Cuvette)

 

Software (integrated)

Operating systemCustom Linux based OS
PC software requirementsWindows® 7 und 10, 64 bit

Does the EzDrop 1000C require a PC to operate?
EzDrop 1000C has a 7-inch high resolution touch panel from which operation can be easily carried out. It can also be operated and controlled from a PC.

What types of cuvettes can be used with the EzDrop 1000C?
Cuvette specification:
Width: 12.5 mm, Length: 12.5 mm, Height:45 mm, detection beam hight: 8.5 mm.
Check the material of the cuvette, if you would like to measure the UV range (190-340 nm), only a quartz cuvette or a UV-grade plastic cuvette can be used.
Each EzDrop 1000C comes with a 10 mm quartz cuvette.

How much volume is needed in the cuvette to make a measurement?
The beam hight of the  cuvette holder in the EzDrop 1000C is 8.5 mm. Ensure that the sample in the cuvette extends beyond 8.5 mm in hight; the volume will vary depending on the cuvette´s specification.
For details, refer to the cuvette supplier´s instruction guide.

How do I clean a sample window that has dried sample residue on it?
Try distilled water. If this is not successful, use 70% ethanol on lint free paper. Wipe the sample window gently. Dry it with fresh lint-free paper. For more detail refer to the operation manual.

What is the lifetime of Xenon lamp?
The lifetime of the Xenon lamp is approximately 10^9 flashes, it can last for 10 years or more. Xenon lamp are very stable and the flash intensity changes very little over the lifetime of the lamp.

How long does the Nano hydrophobic coating on the quartz windows last?
At last 1 year. EzDrop 1000C has a maintenance kit (Cat No. T01-000182-00) to prolong the lifetime of the hydrophobic coating.